20kg Cast Iron Bar Handle Test Mass – Class 3
Part Number: CIC3-K20
Built for durability and reliability, the CIC3-K20 is a 20kg cast iron Class 3 bar handle weight engineered for routine calibration and industrial scale verification. With a sturdy bar handle design and precision casting, it is ideal for environments requiring frequent and practical weight handling.
🔍 Key Features & Benefits
- Nominal Weight: 20 kilograms
- Design: Bar handle for safe and ergonomic lifting and transport
- Material: Heavy-duty cast iron with corrosion-resistant finish
- Tolerance: ±2.0g
- OIML Classification: Class 3 – Suitable for calibrating commercial and industrial scales with moderate resolution
- Applications: Warehouse, production, quality control, and field service calibration
🧪 Optional QA Calibration Service
To ensure measurement accuracy and traceability, QA Calibration can be added at time of purchase. This optional service provides documentation and assurance of conformance to calibration standards. Please allow up to 5 business days for certification processing.
The CIC3-K20 is a robust and economical choice for professionals needing reliable calibration masses in a high-use industrial setting. Ensure consistent scale accuracy with this dependable Class 3 test weight.
Using QA Certified Test Weights for In-House Calibration Checks
To maintain accurate and reliable measurements, it is recommended that customers perform regular verification of their weighing equipment using certified test weights. This practice allows users to identify potential calibration drift between scheduled service intervals and ensures the ongoing precision of weighing systems.
Purchasing certified test weights provides a convenient and effective means for conducting in-house calibration checks. These weights should be selected based on the scale’s maximum capacity and required resolution. All test weights supplied by Triton are traceable to national measurement standards and are issued with calibration certificates to support quality assurance and audit compliance.
Recommended Procedure:
- Ensure the scale is on a level, stable surface and properly zeroed.
- Place the certified test weight on the center of the weighing platform.
- Record the displayed weight and compare it to the known value of the test weight.
- If the deviation exceeds the allowable tolerance, remove the scale from service and arrange for recalibration.
- Log all verification results for traceability and compliance purposes.
Routine use of certified test weights helps maintain measurement integrity and supports compliance with industry standards such as ISO 9001, HACCP, and GLP/GMP where applicable